Jump to content

Mads

Members
  • Posts

    366
  • Joined

  • Last visited

  • Days Won

    21

Mads last won the day on August 16

Mads had the most liked content!

About Mads

  • Birthday 12/01/1975

Profile Information

  • Gender
    Male
  • Location
    Bergen, Norway
  • Interests
    Trail running, skiing, fly fishing, science fiction, food and travel.

LabVIEW Information

  • Version
    LabVIEW 2020
  • Since
    1997

Contact Methods

Recent Profile Visitors

The recent visitors block is disabled and is not being shown to other users.

Mads's Achievements

Rookie

Rookie (2/14)

  • Reacting Well Rare
  • First Post Rare
  • Collaborator Rare
  • Dedicated Rare
  • Week One Done Rare

Recent Badges

65

Reputation

  1. In the example I posted earlier there is a tab at the bottom which (seemingly) scales nicely. It is a trick though...I described it a long time ago here. Maybe someday we get this implemented.... Or better yet, anchoring.
  2. Awesome as in unusual but cool / exploring new concepts in user interface design (often irritating the user in the long run though...) - or as in user friendly and nice (non-LabVIEW-typical?) looking? I too would love to see examples of what others have made. I can share some of the designs I have made myself or have influenced, that I think at least fit the user friendly but not typically LabVIEW-looking bill: We make monitoring systems and use LabVIEW to develop everything that runs on PCs and PACs. I have always put a lot of effort into making the user interfaces as nice, intuitive and recognizable as regular Windows applications (helps with the intuitiveness) as possible in G...The system typically consists of a headless service/embedded server (sbRIO inside a subsea device e.g.) gathering, analyzing and logging data from sensors, and exposing it to control systems or data historians through Modbus, OPC or CANOpen. Users can connect remotely to configure the server and view and analyze the results using application specific clients (through TCP, serial or CANOpen (Cia309.3 transparent links)). Here is a screenshot of the main window of one of the clients (this one is used for corrosion-erosion monitoring): We also make many of the tools we need to do tests and troubleshooting on these sensors and systems, and some of them we have chosen to release as shareware (CAN Monitor on the NI Tools Network e.g.). Here is an example of the Modbus Test Master: We do not earn much from the shareware, but since we want these tools ourselves anyway, and need to maintain them, having them out there at least do not cost us much. It also puts a bit of extra pressure on us to make even (some of) the internal tools look and work well.
  3. I agree with you, and would normally stick to just a File\Exit menu option and window close. As long as extra ways of closing the application do not harm (read: confuse) more than they help I would not argue too much against them though. It depends mostly on the type of application. Different people prefer different solutions (the true lesson in the 80/20 rule). A File>Exit (Ctrl+Q) menu option and the window close covers the standard methods, so they should always be there, but if the application might end up running on a computer with a touch screen, and access to an exit is desirable, having a button would be nice too - and not weird. (Assuming it is made distinct from other buttons that just stop sub-processes (as ShaunR mentioned earlier)). I am lucky to normally have the final say in such matters myself😃. PS. We used to have a financial system that did not offer any save options. Instead you were always required to select print, and then Print to file (💩!). I would love to meet and have a chat with the people who designed and approved that solution😉. (This was even before printing to pdf e.g. became a concept, which at least makes it partially recognizable as an *optional* route today).
  4. I have to agree with the requirement (although having an OK and a Cancel would suffice unless you also want to apply prior to closing the window). Stick to the norms established by the OS, familiarity is key. If I regret entering the dialog and/changing something having OK/Cancel there will allow me to recognise what to do in a fraction of a second. If this particular designer though it obvious that closing the window would revert...eh, I mean save the...hmmm. Do I need to change back the changes I did, or is it enough to close the window, I wonder....🤔
  5. Hi, My guess is that you have set the Get Image method to use 16 bit colors. Use 24 bit and it will show up correct (I recreated and solved the issue just now by generating a front panel image of a screenshot of your original colors).
  6. What is the cost you are worried about? Is it speed, memory, or both? If it is speed - is the example representative? In that case I would calculate the window function within the inner loop and just output the result, not generate an array of all the subwindows and *then* apply the actual window calculation. The loop(s) can also be configured for parallelism to speed things up significantly (assuming you have more than one core at least)... More speed, less memory. If none of that reduces the cost enough you will probably have to do the processing in external (non-G) code.
  7. If only NI could implement my second most popular idea on the idea exchange 😉 (please vote!). (Hopefully they can squeeze it in while working on my most popular idea/rant which partially covers this one anyway).
  8. What error code does the Modbus VIs return, if any? Can you try it towards a local Modbus TCP server/simulator, does that work better? Can you temporarily disable the firewall? I used your code as it was, just changed the IP, and it connected immediately to my local Modbus server and was able to read the registers, I also changed the addresses out of that server's range to verify that I got the expected exception response/error I showed earlier. If you can install Wireshark I would use that to sniff and compare the traffic from the LabVIEW code with the one you generate with Qmodmaster (post the Modbus traffic here and we can decode, Wireshark can help you as it is able to decode Modbus too).
  9. A couple of notes: a) The code you have will work for reading *Holding* registers (not input) from a slave/server on the given IP and port, but only if that server ignores the unit id or happens to have unit id=1. If it requires a different unit id (many Modbus TCP servers expect it to be 0), right click on the modbus master wire and insert the "Set Unit ID"-function, then wire the correct id to that. b) Assuming the unit id is not preventing the server from bothering to answer you should still get replies from the device even if it does not have any holding registers available, but the replies will then be exception responses and the functions you use will then return an error looking like this (create an indicator within the loop to see it imemdiately): Error -389112 occurred at an unidentified location Possible reason(s): LabVIEW: (Hex 0xFFFA1008) The Modbus slave does not accept the data address contained in the query. Function 3 c) Why two Modbus master connections to the same device? It might support it, but during troubleshooting I would remove the second loop and focus on getting one working.
  10. I'll leave the logic to others for now, but here are a couple of initial observations: 1) If these tables can get larger you might consider making the table cell formatting more efficient: Defer front panel updates (updating table GUI for each step is extremely slow) then format the table cells and headers in one go before you reenable front panel updates. If all rows of all columns are to have the same format anyway (empty ones can get the same as they are empty anyhow) you might as well set the active cell to -2,-2 and apply to all. If the headers are to be different you can then apply a reformatting of all of them afterwards using a -1, -2 combo (kudo this idea to avoid that last step in the future...). 2) The OpenG toolkit has "Remove Duplicates" and "Sort 2D array" functions you could use to get the unique and sorted week and models. If you have LabVIEW 2020 the same functions are available as VIMs on the array palette. 3) Avoid the repeated searches, e.g. by running an auto-index-driven set of for loops and simple equal or not checks instead. As others have mentioned already using a feedback node / shift register to keep track of whether to insert a 0% or not is likely to be part of the solution. Personally I would probably be lazy and just initialize the whole table with 0% and leave it at that 😄, but that might be misleading here...and not as pretty. Working from the bottom up filling in 0% on empty cells after the last entry has been found (state kept in feedback/shift reg) is another idea. The fact that you want to fill APR W3 all the way down to the end of APR W2 makes it necessary to include the state (end index) from the previous column as well though, not just whether or not there is an entry lower in the same column.
  11. I have not seen that. It is not the NI Tools Network your thinking of, is it? We did put one product there as it relies on NI hardware, but otherwise it feels more right for libraries and coding tools than generic applications...
  12. I recently needed to figure out a good way to get a piece of shareware (developed i LabVIEW) distributed to the various popular download sites. It does not really matter which programming language or tool(s) were used in the development so I did not search for help on this here or at ni.com. One good source I found was this https://www.fatbit.com/fab/promote-software-app-online/ It did strike me as a bit curious though that we never really discuss such matters. Most applications developed in LabVIEW are probably not generic enough to be released through such channels, but sometimes they are, and unless you have your own sales and marketing people that are familiar with software you might need to do the job yourself. Perhaps some of you have been through this with your products already and have some tips and tricks to share? One generic product we have *not* released this way yet uses the OPC UA API. The distribution license requirement NI puts on it makes it too cumbersome to handle 😒 One of the things I discovered from the mentioned web site was the use of PAD-files. I created an account at AppVisor and started working on a file for the shareware we were to release. You can edit, save and publish the PAD file, but very few download sites automatically discover your software that way so I soon came to realize that I needed to pay AppVisor a fee for them to submit the software to the download sites. I did try to submit to my favourite download site myself first (download.com), but they never published it on their site. The AppVisor submission service came in different price categories and as I did not expect the software to generate that much sales I chose the cheapest option for now which only gives you 4 submissions per year to 100 sites, but it is a starting point. One thing I soon learned was that the download links should be made permanent. At first I used links that contained the version and build number , and that got cumbersome very quickly as we soon needed to submit a few bug fixes and updates... I also had to create program descriptions of various standardized lengths, provide links to icons, screen captures etc. Having a video or two showing some of the features is probably a good idea as well. Reading tips and tricks from other sites I started mentioning and/or linking to the product on social media, but tried to do it rather discreetly as I personally hate obvious marketing🤢. I still got marked very quickly as "spam" on one site - even though I had only mentioned the product in a relevant setting, and along with similar competing products 😦. It makes me appreciate the work our sales people normally do for us. They are too busy selling our main products so I have not bothered them much with this.
  13. I have seen this a couple of times in my applications and a fix then has always been to either remove all but English, or support all languages on the run-time languages page in the application builder (perhaps any change there is what is really doing the trick, I have not explored that as I have always just moved on when the error disappeared).
  14. Just to tie off this thread: The reason for the glitch was that the wiring allowed for a minuscule race-condition 😒. One minor adjustment was needed to ensure everything was *only* decided by the enqueue scheduling. So no worries about that anymore 😀
  15. Nice discussion, thanks for the link. The quoted statement seems to contradict my observations yes. I have not checked if the misbehavior was absent in any of my earlier LabVIEW installations yet...I am working in Windows LabVIEW 2020 SP1 at the moment.
×
×
  • Create New...

Important Information

By using this site, you agree to our Terms of Use.